# Capnography | ER Basics

Web page: https://erbasics.com/sheets/capnography.html

Cheat Sheet

# Capnography

Normal values, waveform patterns, CPR and ROSC, tube checks and sedation monitoring

Reviewed by [Alex G, BSN, RN, CEN, CCRN, CMC, NREMT-P](https://icubasics.com/about.html) · Checked against current sources October 2026

[Practice 15 Capnography questions free](https://erbasics.com/app/?from=/sheets/capnography.html#/topic/capnography)

[![Page 1 of the Capnography PDF, a printed page in columns with sections on Normal values and phases, Patterns and causes, CPR quality and ROSC (AHA 2025), Confirming tube placement and Sedation and opioids](https://erbasics.com/sheets/thumbs/capnography.webp) Enlarge](https://erbasics.com/sheets/thumbs/capnography-lg.webp)

**One-page PDF** Letter landscape, ready to print or keep with your notes

[View PDF](https://erbasics.com/sheets/capnography.pdf) [Download](https://erbasics.com/sheets/capnography.pdf)

## Normal values and phases

EtCO₂: the CO₂ at the end of each exhaled breath. Typical adult values for exam study.

| Item | Normal |
| --- | --- |
| EtCO₂ | 35–45 mm Hg |
| PaCO₂ minus EtCO₂ | About 2–5 mm Hg |
| Baseline | Returns to zero |
| Shape | Rectangular: steep rise, plateau flat or gently rising |

| Phase | What it shows |
| --- | --- |
| I | Flat baseline: dead-space gas, no CO₂ |
| II | Steep upstroke as alveolar gas arrives |
| III | Alveolar plateau; EtCO₂ read at its end |
| Downstroke | Inspiration: falls to zero (phase 0; some texts say IV) |

*Normal: zero baseline, steep upstroke, gently rising plateau peaking near 38 mm Hg, sharp drop on inspiration.*

A wider PaCO₂–EtCO₂ gap means more dead space: low cardiac output, hypovolemia, [PE](https://erbasics.com/sheets/pulmonary-embolism.html), emphysema. The gap varies in sick lungs, so match EtCO₂ to an [ABG](https://erbasics.com/sheets/abg-steps.html) before trending it.

## Patterns and causes

*Shark fin: sloping upstroke, no plateau. Bronchospasm or a kinked or plugged tube.*

*Sudden loss: normal breaths, then flat at zero. Disconnection, dislodged tube, complete obstruction or apnea until proven otherwise.*

| Pattern | Think |
| --- | --- |
| Sudden drop to zero | Disconnection, extubation or dislodged tube, complete obstruction, apnea |
| Sudden drop, not to zero | Leak, partial obstruction, abrupt fall in cardiac output (PE, arrest) |
| Sudden rise | ROSC, IV sodium bicarbonate, tourniquet release |
| Gradual rise | Hypoventilation, fever; malignant hyperthermia if it climbs despite more ventilation |
| Gradual fall | Hyperventilation, hypothermia, falling cardiac output |
| Shark fin | Bronchospasm (asthma, COPD), kinked or plugged tube |
| Baseline above zero | Rebreathing, exhausted CO₂ absorber, contaminated sensor |
| Notch in plateau | Curare cleft: breathing effort as paralysis wears off |

**Safety** When the trace suddenly flattens, assess the patient before the monitor: tube, circuit, breathing, pulse.

**Exam tip** Exams pair shark fin with bronchospasm, sudden zero with a disconnected or dislodged tube, and gradual rise with hypoventilation.

Try one

Capnography practice question 1 of 3

During a PEA arrest, an intubated patient is log-rolled onto a backboard. Before the move, waveform capnography showed a steady trace with EtCO₂ 18 mm Hg. Now the trace is flat at 0, no breath sounds are heard on either side, and gurgling is heard over the epigastrium. Compressions are still high quality. Which action is MOST appropriate?

A. Give epinephrine now, since a flat EtCO₂ means no cardiac output
B. Advance the tube 2 cm and listen again at the next rhythm check
C. Remove the tube, ventilate by bag-mask, and replace the airway
D. Raise the ventilation rate to bring the EtCO₂ back up

**Show the answer and rationale**

**Answer: C**

- **A.** Falling output lowers EtCO₂ gradually as compressions tire. A sudden drop to zero with high-quality compressions, absent breath sounds and epigastric sounds points to the airway, not the circulation.
- **B.** Advancing helps a tube that has backed up toward the cords, but epigastric gurgling with no breath sounds means the tip is in the esophagus; pushing it deeper keeps the patient unventilated for another 2 minutes.
- **C.** Correct. The findings show a tube that has slipped into the esophagus. Remove it, oxygenate by bag-mask or a supraglottic airway, then place a new airway and confirm it with waveform capnography.
- **D.** Faster breaths cannot raise the EtCO₂ when the tube no longer sits in the trachea, and overventilation in arrest raises intrathoracic pressure and lowers venous return.

Hypoxia is one of the Hs, and a displaced tube is a common cause of it during resuscitation. Waveform capnography is recommended both to confirm tube placement and to monitor it continuously, and the position is rechecked after every move or transport. A gradual fall in EtCO₂ suggests tiring compressions or falling output. A sudden loss of the waveform while compressions stay high quality, with no breath sounds and gurgling over the stomach, means esophageal displacement. The fix is to remove the tube, oxygenate with a bag-mask device or supraglottic airway, and replace the airway with capnography confirmation, keeping any compression pause under about 5 seconds.

Sources: [2025 Korean CPR Guidelines: Adult ALS](https://pmc.ncbi.nlm.nih.gov/articles/PMC13277415/); [AHA 2025 Part 9: Adult ALS](https://cpr.heart.org/en/resuscitation-science/cpr-and-ecc-guidelines/adult-advanced-life-support); [AHA ACLS Provider Manual (2025)](https://shopcpr.heart.org/acls-provider-manual)

[Practice 14 more questions free](https://erbasics.com/app/?from=/sheets/capnography.html#/topic/capnography) Rationales for every answer. No download needed.

## CPR quality and ROSC (AHA 2025)

During CPR, EtCO₂ mainly reflects the blood flow compressions generate. The 2025 algorithm: if EtCO₂ is low or falling, reassess CPR quality.

| EtCO₂ during CPR | Suggests |
| --- | --- |
| Under 10 mm Hg | Little flow: check depth, rate, recoil, pauses; swap compressors |
| 10 or more, ideally 20 or more | Linked to adequate compressions and more ROSC |
| Abrupt, sustained rise | Possible ROSC (a jump over 10 mm Hg is suggestive; smaller rises occur); confirm with pulse or arterial line |
| 10 or less after 20 min of ALS, intubated | One factor in deciding to stop, never the only one |

Ventilation rate, [epinephrine](https://erbasics.com/drugs/epinephrine.html) and [sodium bicarbonate](https://erbasics.com/drugs/sodium-bicarbonate.html) also shift EtCO₂. With an advanced airway, give 1 breath every 6 seconds.

**Exam tip** Low or falling EtCO₂: fix compressions, check the ventilation rate, confirm the tube. Abrupt sustained rise: check for ROSC. Older materials quote about 40 mm Hg; the 2025 guideline names no cutoff.

**Safety** In patients who are not intubated, no EtCO₂ cutoff should be used to stop resuscitation (AHA 2025, Class 3: Harm).

## Confirming tube placement

- In cardiac arrest, waveform capnography plus clinical assessment is the most reliable way to confirm and monitor the tube (AHA 2025, Class 1). ASA standards: verify every tracheal tube or supraglottic airway with exhaled CO₂.
- Look for sustained CO₂: peaks that rise and fall with each breath and hold or grow over at least 7 breaths. CO₂ that fades suggests the esophagus.
- In arrest, CO₂ can be low, but good compressions should not leave a completely flat trace.
- Colorimetric detector: purple = no CO₂, yellow = CO₂; check the device's key. Arrest can give a false negative; gastric acid or drugs, a false positive that does not change breath to breath.
- Keep monitoring; sudden loss may be the first sign of dislodgement.

**Safety** No trace = wrong place, even in arrest. Consensus guidance makes removing the tube the default; if it stays, recheck with video laryngoscopy, bronchoscopy or ultrasound. Breath sounds cannot rule out the esophagus.

## Sedation and opioids

ASA standards (amended 2025): during moderate or deep sedation, observe ventilation continually and make every effort to monitor exhaled CO₂.

- Capnography shows hypoventilation and apnea before SpO₂ falls, especially when oxygen keeps saturation up (procedural sedation, opioid PCA). It is linked to fewer desaturations.
- Slow breathing (typical of opioids): high EtCO₂, tall, wide waveform.
- Shallow breathing (typical of sedative-hypnotics): small waveform, normal or low number, while PaCO₂ rises. A low reading is not reassuring.
- Shrinking or absent waveform while the chest moves: obstruction; open the airway.
- Flat line, no effort: apnea; stimulate and ventilate.
- Oxygen dilutes nasal cannula samples, so readings run low; compare trend and shape with baseline.

[Practice Capnography questions free in your browser](https://erbasics.com/app/?from=/sheets/capnography.html#/topic/capnography) 15 exam-style questions with rationales. No account or download needed.

## In depth

The one-page summary above is what the PDF prints. These sections go further.

### CO₂ production, transport and elimination

A resting adult produces about 200 mL of CO₂ a minute (VCO₂) while consuming about 250 mL of O₂, a respiratory quotient near 0.8. The quotient is 1.0 when burning carbohydrate and 0.7 for fat, so overfeeding with carbohydrate raises VCO₂ and the ventilation needed to clear it. Fever raises metabolic rate by roughly 10–13% per °C; shivering, seizures, [sepsis](https://erbasics.com/sheets/sepsis-bundle.html), [burns](https://erbasics.com/sheets/burns.html) and malignant hyperthermia raise it further, while [hypothermia](https://erbasics.com/sheets/environmental-emergencies.html), sedation and paralysis lower it.

CO₂ is about 20 times more soluble than O₂. Of the CO₂ carried from tissue to lung, roughly 70% travels as bicarbonate (carbonic anhydrase in red cells, with the chloride shift), 20–25% as carbamino compounds on hemoglobin and 5–10% dissolved. Deoxygenated hemoglobin binds more CO₂ (the Haldane effect), so oxygenation in the lung helps unload it. Mixed venous PCO₂ is about 46 mm Hg against an arterial 40.

Elimination follows the alveolar ventilation equation: PaCO₂ = 0.863 × VCO₂ ÷ alveolar ventilation, where alveolar ventilation = (tidal volume − dead space) × rate. Halve alveolar ventilation and PaCO₂ doubles. The body holds large CO₂ stores, so after sudden apnea PaCO₂ is commonly quoted as rising about 10–15 mm Hg in the first minute and 3–5 mm Hg per minute after that, while SpO₂ on supplemental oxygen can stay normal for minutes.

*Every EtCO₂ change comes from one of three places.*

| Determinant | EtCO₂ rises with | EtCO₂ falls with |
| --- | --- | --- |
| Production (metabolism) | Fever, sepsis, MH, bicarbonate | Hypothermia, sedation, paralysis |
| Delivery (pulmonary blood flow) | ROSC, rising cardiac output | Arrest, shock, PE, hemorrhage |
| Elimination (alveolar ventilation) | Hypoventilation, obstruction | Hyperventilation, apnea, disconnection |

**Exam tip** When ventilation is held constant, as on a controlled ventilator or during CPR, EtCO₂ tracks pulmonary blood flow. That is why it falls first in arrest, PE and hemorrhage and rises first at ROSC.

### How capnographs measure

Infrared capnographs pass light at about 4.26 µm, the CO₂ absorption peak, through the gas; absorption is proportional to the CO₂ partial pressure. Capnometry gives only the number; capnography adds the waveform, which is what shows obstruction, rebreathing and artifact. Values appear in mm Hg, kPa (1 kPa ≈ 7.5 mm Hg) or percent (5% ≈ 38 mm Hg at sea level). Monitors show a fast sweep for shape and a slow trend for minutes to hours.

| Feature | Mainstream | Sidestream |
| --- | --- | --- |
| Where gas is read | Sensor on an adapter at the airway | Gas pulled through tubing to the monitor |
| Sample flow | None | 50–250 mL/min (microstream about 50) |
| Delay | Almost none | Transit delay of seconds; longer line, longer delay |
| Patients | Mainly intubated | Intubated or not (nasal and oral cannulas) |
| Drawbacks | Weight and adapter dead space at the tube | Water and secretions block the line |
| Small tidal volumes, fast rates | Accurate | May underread; sample can entrain fresh gas |

Colorimetric detectors use pH-sensitive paper that changes color with CO₂. They confirm that CO₂ is present but give no waveform and no number, they cannot monitor continuously, and they lose accuracy when wet. They remain useful where waveform capnography is not available and, in some protocols, to detect a small-bore feeding tube entering the airway during insertion, which still needs radiographic confirmation.

### The normal capnogram: phases, angles and slope

Phase I is the start of exhalation: gas from the airways and equipment, which held no gas exchange, so CO₂ stays at zero. Phase II is the steep rise as alveolar gas mixes with dead-space gas. Phase III is the alveolar plateau, which rises slightly because alveoli with low ventilation-to-perfusion ratios and higher CO₂ empty last. EtCO₂ is the value at the end of phase III. Inspiration (phase 0, or IV in some texts) brings fresh gas and a sharp fall to zero.

*Angles from the time-capnogram teaching on capnography.com (Kodali).*

| Feature | Normal | What changes it |
| --- | --- | --- |
| Alpha angle (phase II to III) | About 100–110° | Widens with bronchospasm, V/Q mismatch |
| Beta angle (phase III to downstroke) | About 90° | Widens with rebreathing on inspiration |
| Phase III slope | Flat or gently rising | Steeper in COPD and asthma |
| Terminal upswing (phase IV′) | Absent | Pregnancy, obesity: EtCO₂ may exceed PaCO₂ |

Shape tells you about the airway; height tells you about ventilation, perfusion and production. A tall wide waveform is slow breathing with CO₂ retention. Short narrow waves are fast or shallow breaths. A sloping upstroke without a plateau means alveoli are emptying unevenly through narrowed airways, so the number at the end of that slope underestimates alveolar CO₂.

### PaCO₂–EtCO₂ gradient, dead space and volumetric capnography

Physiologic dead space is anatomic dead space (about 2 mL/kg, roughly 150 mL in an adult) plus alveolar dead space, alveoli that are ventilated but not perfused. Gas from unperfused alveoli contains almost no CO₂ and dilutes the gas from perfused alveoli, so EtCO₂ falls below PaCO₂. The normal gap is about 2–5 mm Hg; it widens with low cardiac output, hypovolemia, PE, emphysema, high PEEP that overdistends alveoli, and age. It can be zero or negative in pregnancy and with large tidal volumes at slow rates, where the late-emptying alveoli carry CO₂ above the arterial value.

The Bohr equation gives the dead-space fraction: VD/VT = (PACO₂ − PĒCO₂) ÷ PACO₂, where PĒCO₂ is mixed expired CO₂ collected over a minute or measured by volumetric capnography. Enghoff substituted PaCO₂ for the alveolar value, VD/VT = (PaCO₂ − PĒCO₂) ÷ PaCO₂, which is what bedside calculation uses. Because PaCO₂ also carries the effect of shunt and V/Q mismatch, Enghoff dead space is larger than true Bohr dead space. Normal VD/VT is about 0.3 breathing spontaneously; equipment adds to it on the ventilator.

In early [ARDS](https://erbasics.com/sheets/ards.html), Nuckton 2002 (NEJM, 179 patients) found a mean dead-space fraction of 0.58, and each 0.05 increase raised the odds of death by about 45%, independent of oxygenation. A simpler bedside surrogate is the alveolar dead-space fraction, (PaCO₂ − EtCO₂) ÷ PaCO₂. Some PE studies combined a normal value with a negative D-dimer to exclude PE, but no guideline uses it on its own.

Volumetric capnography plots CO₂ against exhaled volume rather than time, using a mainstream sensor with a flow sensor. Phase I volume is anatomic plus equipment dead space (Fowler's method measures it at the midpoint of phase II). The area under the curve is the CO₂ eliminated per breath; summed over a minute it gives VCO₂. From these the monitor reports VD/VT, alveolar tidal volume and VCO₂ breath by breath.

- PEEP titration: dead space is lowest near the PEEP with best compliance; rising VD/VT as PEEP climbs suggests overdistension.
- Weaning: VCO₂ that climbs during a spontaneous breathing trial means rising work of breathing.
- Energy expenditure: VCO₂ (mL/min) × 8.19 estimates kcal/day when indirect calorimetry is unavailable (Stapel 2015).
- Perfusion: a sudden fall in VCO₂ at constant ventilation means less pulmonary blood flow.

**Safety** Never assume a fixed 5 mm Hg gap. Establish it with an ABG, and recheck after changes in cardiac output, PEEP, position or lung disease, because the gap moves with them.

### Abnormal waveforms, pattern by pattern

| Pattern | Mechanism or cause | Response |
| --- | --- | --- |
| Shark fin, no plateau | Uneven emptying: bronchospasm, COPD, kinked tube | Suction, bronchodilator, check tube |
| Flat line at zero | Disconnection, dislodged tube, apnea, obstruction | Patient first: tube, circuit, breathing, pulse |
| Exponential fall over 1–2 min | Sudden loss of pulmonary flow: arrest, PE, hemorrhage | Pulse and blood pressure now |
| Lower plateau, irregular | Cuff leak, partial extubation, sampling-line leak | Check cuff, tube depth, connectors |
| Baseline above zero | Rebreathing, exhausted absorber, contaminated sensor | Check circuit and valves; zero or replace sensor |
| Slanted downstroke | Inspiratory valve fault, slow sampling at fast rates | Check valve; compare with mainstream |
| Cleft in plateau | Curare cleft: diaphragm effort as blockade wears off | Assess train-of-four and sedation |
| Ripples at end of plateau | Cardiogenic oscillations from heartbeats at slow rates | None; benign |
| Small waves that fade | Esophageal tube after bag-mask air in stomach | Treat as esophageal; remove or confirm |
| Small waves between breaths | Spontaneous effort or asynchrony on the ventilator | Assess sedation, trigger, settings |
| Tall wide waves, rising trend | Hypoventilation or rising production | Assess drive, airway, temperature |
| Short narrow waves, falling trend | Hyperventilation, anxiety, pain, acidosis | Look for the cause: pain, acidosis, sepsis |

Separate a fall to zero from a fall to a low value. A drop to zero in one breath is mechanical (no gas reaching the sensor) until proven otherwise. A fall over several breaths to a low but present value, with an unchanged ventilator, is perfusion. A sustained rise despite more minute ventilation is production, and in an anesthetized patient with rising temperature, tachycardia and rigidity that is malignant hyperthermia until proven otherwise.

**Exam tip** In bronchospasm the reported EtCO₂ is falsely reassuring: the plateau never forms, so the true alveolar value is higher. Watch the shape improve with treatment, and get an ABG.

### Tube confirmation, cardiac arrest and prognosis

Tracheal gas carries CO₂ breath after breath as long as there is any pulmonary blood flow. Esophageal gas can contain CO₂ after bag-mask ventilation or carbonated drinks, but it washes out over a few breaths, which is why sustained CO₂ over at least 7 breaths is the test (PUMA 2022). Capnography confirms that the tube ventilates perfused lung; it does not detect a mainstem intubation, and a tube whose tip sits just above the cords can still show a waveform. Depth, bilateral chest rise and a chest film still matter.

During CPR, cardiac output is roughly a quarter to a third of normal and EtCO₂ follows it closely at a fixed ventilation rate. Overventilation lowers EtCO₂ and a slow rate raises it, so the number only reflects compressions when ventilation is steady at 1 breath every 6 seconds (AHA 2025). Sodium bicarbonate releases CO₂ (50 mEq can generate up to about 1 L), giving a transient rise that is not ROSC. Epinephrine can lower EtCO₂ for a few minutes despite raising coronary perfusion pressure.

Levine 1997 (NEJM, 150 out-of-hospital arrests with PEA) found that no patient with EtCO₂ of 10 mm Hg or less after 20 minutes of ALS survived to admission, which is the origin of the termination factor in the AHA guideline. Later cohorts have reported survivors below that value, and the number is confounded by ventilation rate, drugs, airway position and the cause of arrest. Asphyxial arrests start with high EtCO₂ as stored CO₂ washes out, so early values say little about prognosis.

**Safety** AHA 2025 makes low EtCO₂ after 20 minutes a factor to consider in intubated patients only (2b), never the sole reason to stop, and calls using any cutoff in non-intubated patients harmful. Supraglottic airway readings are less reliable because of leaks.

### Sedation, opioids, nasal capnography and transport

Pulse oximetry measures oxygenation; capnography measures ventilation. With supplemental oxygen, SpO₂ can stay above 95% for minutes while PaCO₂ climbs, so desaturation is a late sign of respiratory depression. The Joint Commission (Sentinel Event Alert 49, 2012) called for assessing both oxygenation and ventilation in patients on opioids, and capnography is the practical way to watch ventilation on the ward and during procedural sedation.

Nasal and oral-nasal sampling cannulas draw exhaled gas from the nares and mouth. Room air and oxygen flow dilute the sample, so the number runs low and is best used as a trend with the waveform and respiratory rate. Capnography gives a more reliable respiratory rate than chest impedance and detects apnea within seconds; set the no-breath alarm to the local policy, usually 20–30 seconds. Mouth breathers need an oral scoop, and high-flow oxygen can wash out the sample entirely.

- Early opioid effect: rate falls, waveform widens and EtCO₂ rises before SpO₂ changes.
- Sedative-hypnotics: shallow breaths, small waveform and a normal or low EtCO₂ while PaCO₂ rises.
- Partial obstruction: smaller, irregular waves; snoring; chest moving. Reposition, jaw thrust or airway adjunct.
- Apnea: no waveform, no effort. Stimulate, ventilate, consider [naloxone](https://erbasics.com/drugs/naloxone.html) or reversal.

In transport, disconnection and dislodgement are most likely during moves between bed, stretcher and aircraft, so keep the waveform visible and recheck after every move. Cold and humidity changes cause condensation in the line or on the sensor window. Monitors that report percent CO₂ depend on barometric pressure; most transport monitors report mm Hg with automatic compensation, but confirm this for flight. For head injury, target normocapnia and confirm with an ABG, because the gap widens with hypovolemia and chest trauma (Brain Trauma Foundation 2016 advises against prophylactic hyperventilation to PaCO₂ of 25 or less).

### Ventilated patients, PE, sepsis and perfusion

On the ventilator, continuous EtCO₂ gives early warning of disconnection, tube migration, obstruction, rising dead space and changes in metabolism, but it replaces an ABG only after the gap is known and stable. A widening gap at constant settings points to a fall in cardiac output, a PE, or overdistension from PEEP or prone positioning. EtCO₂ that rises during a spontaneous breathing trial, with a faster rate, suggests the patient is tiring.

In PE, clot creates alveolar dead space: EtCO₂ falls and the PaCO₂–EtCO₂ gap widens while minute ventilation is unchanged or higher. Massive PE can cause an abrupt fall that mimics a disconnection but leaves a waveform. EtCO₂ rising after thrombolysis suggests reperfusion. Capnography cannot diagnose PE on its own; it prompts imaging.

In sepsis and other metabolic acidosis, compensatory hyperventilation lowers PaCO₂ and poor perfusion adds dead space, so EtCO₂ falls as lactate rises. Hunter 2013 (Am J Emerg Med) found EtCO₂ inversely related to lactate and mortality in suspected sepsis, and some EMS sepsis screens use 25 mm Hg or less as a trigger. EtCO₂ does not replace a lactate level. The same mechanism explains low EtCO₂ in [DKA](https://erbasics.com/sheets/dka-hhs.html).

Under constant ventilation, a rise in EtCO₂ during passive leg raising reflects a rise in cardiac output. Monnet 2013 (Intensive Care Med) found a rise of 5% or more predicted fluid responsiveness with high specificity but modest sensitivity. Spontaneous breathing and changing [ventilator settings](https://erbasics.com/sheets/vent-modes.html) invalidate the test.

### Troubleshooting

| Problem | Likely cause | Action |
| --- | --- | --- |
| No waveform, patient breathing | Kinked or blocked line, line not attached | Check line; replace if wet or blocked |
| Occlusion alarm | Water or secretions in line or filter | Empty or replace water trap; new line |
| Reading lower than expected | Dilution by O₂ flow, leak, fast rate on sidestream | Fix connectors; compare with ABG |
| Baseline above zero | Contaminated sensor, missed zero, rebreathing | Zero the sensor; check circuit and valves |
| Flat line for a few seconds | Automatic zeroing on a sidestream monitor | Wait; it returns by itself |
| Mainstream reading drifts | Window fogged or adapter changed | Clean or change adapter; zero again |
| Values off on known gas | Calibration out of date | Remove from use; send to biomedical |

- Keep the sampling port above the tube so condensate drains away from it, and follow the maker's guidance on where the port sits relative to the HME filter.
- Use the same cannula type for trending; switching from nasal to oral-nasal or from sidestream to mainstream changes the baseline.
- High oxygen or nitrous oxide concentrations can bias infrared readings; most current monitors compensate automatically, but some ask for the gas mix.
- Small infants on sidestream monitors can show falsely low values because the sample flow approaches their expiratory flow.

**Safety** Troubleshoot the monitor only after the patient is safe. A flat trace in an intubated patient is a dislodged tube, disconnection or arrest until you have checked all three.

### Exam pearls

- EtCO₂ depends on production, pulmonary blood flow and alveolar ventilation; at constant ventilation it tracks cardiac output.
- Normal EtCO₂ 35–45 mm Hg with a PaCO₂–EtCO₂ gap of about 2–5; a wider gap means more dead space.
- Shark fin with a widened alpha angle is bronchospasm; baseline above zero is rebreathing.
- Sudden zero is a disconnected or dislodged tube or apnea; an exponential fall over a minute or two is lost perfusion.
- In CPR, under 10 mm Hg means fix compressions; an abrupt sustained rise suggests ROSC; bicarbonate gives a transient rise.
- Capnography detects hypoventilation before SpO₂ falls, especially on supplemental oxygen.
- Waveform capnography confirms the tube is in the airway but not its depth; it misses mainstem intubation.
- Curare cleft is a notch in the plateau from returning diaphragm effort.
- Gradual rise despite more ventilation in the OR suggests malignant hyperthermia.
- In head injury, confirm EtCO₂ against an ABG before using it to guide ventilation.

[Practice free in your browser](https://erbasics.com/app/?from=/sheets/capnography.html) 1,900+ exam-style questions with rationales. Free today, no account, no ads. Pro is coming in January 2027. Also free [on Google Play](https://play.google.com/store/apps/details?id=com.alex.erbasics).

## Related cheat sheets

- [**ABG Interpretation in 6 Steps** pH, PaCO₂, HCO₃⁻, compensation, anion gap and oxygenation, with worked examples](https://erbasics.com/sheets/abg-steps.html)
- [**Ventilator Modes and Settings** Modes, lung-protective targets, peak vs plateau, alarms and SBT readiness](https://erbasics.com/sheets/vent-modes.html)
- [**ARDS** Berlin and Global definitions, lung-protective ventilation, PEEP, proning, paralysis, steroids, ECMO and the nursing care](https://erbasics.com/sheets/ards.html)
- [**Pulmonary Embolism** ESC 2019 and AHA/ACC 2026 risk classes, diagnosis, anticoagulation, lysis, catheter therapy and RV support](https://erbasics.com/sheets/pulmonary-embolism.html)
- [**Asthma and COPD Exacerbations** Severity signs, first treatments, oxygen targets, NIV and ventilating obstructive lungs](https://erbasics.com/sheets/asthma-copd.html)
- [**Chest Tubes and Pneumothorax** Tension pneumothorax, the drainage system, bubbling and tidaling, troubleshooting and removal](https://erbasics.com/sheets/chest-tubes.html)

## Calculators for this topic

- [Coronary Perfusion Pressure](https://erbasics.com/calculators/coronary-pp.html)
- [Dead Space Fraction (Bohr)](https://erbasics.com/calculators/bohr-deadspace.html)
- [Shunt Fraction (Qs/Qt)](https://erbasics.com/calculators/shunt.html)

## Sources

1. Wigginton JG, Agarwal S, Bartos JA, Coute RA, et al. Part 9: Adult Advanced Life Support: 2025 American Heart Association Guidelines for Cardiopulmonary Resuscitation and Emergency Cardiovascular Care. Circulation. 2025;152(16 suppl 2):S538–S577. doi:10.1161/CIR.0000000000001376 [Link](https://www.ahajournals.org/doi/full/10.1161/CIR.0000000000001376)
2. Panchal AR, Bartos JA, Cabañas JG, et al. Part 3: Adult Basic and Advanced Life Support: 2020 American Heart Association Guidelines for Cardiopulmonary Resuscitation and Emergency Cardiovascular Care. Circulation. 2020;142(16 suppl 2):S366–S468. doi:10.1161/CIR.0000000000000916 [Link](https://www.ahajournals.org/doi/full/10.1161/CIR.0000000000000916)
3. Chrimes N, Higgs A, Hagberg CA, et al. Preventing unrecognised oesophageal intubation: a consensus guideline from the Project for Universal Management of Airways and international airway societies. Anaesthesia. 2022;77(12):1395–1415. doi:10.1111/anae.15817 [Link](https://pmc.ncbi.nlm.nih.gov/articles/PMC9804892/)
4. American Society of Anesthesiologists. Standards for Basic Anesthetic Monitoring. Committee on Standards and Practice Parameters; last amended October 15, 2025 [Link](https://www.asahq.org/standards-and-practice-parameters/standards-for-basic-anesthetic-monitoring)
5. Kodali BS, Urman RD. Capnography during cardiopulmonary resuscitation: current evidence and future directions. J Emerg Trauma Shock. 2014;7(4):332–340. doi:10.4103/0974-2700.142778 [Link](https://pmc.ncbi.nlm.nih.gov/articles/PMC4231274/)
6. Kodali BS. Components of a time capnogram. Capnography.com [Link](https://www.capnography.com/components-of-a-time-capnogram/)
7. Kodali BS. (a-ET)PCO2 difference and alveolar dead space. Capnography.com [Link](https://www.capnography.com/a-etco2-difference-and-alveolar-dead-space/)
8. Kodali BS. Capnography in intensive care unit. Capnography.com [Link](https://www.capnography.com/canography-in-intensive-care-unit/)
9. Kodali BS. Interpret your capnogram. Capnography.com [Link](https://www.capnography.com/interpret-your-capnogram/)
10. Galvagno SM, Kodali BS. Emergency medicine and prehospital critical care. Capnography.com [Link](https://www.capnography.com/emergency-medicine-and-prehospital-critical-care/)
11. Kodali BS. Chemical method of carbon dioxide measurement. Capnography.com [Link](https://www.capnography.com/chemical-method-of-carbon-dioxide-measurement/)
12. Kodali BS. Royal College of Anaesthetists educational campaign on capnography during CPR ('No Trace = Wrong Place'). Capnography.com [Link](https://www.capnography.com/royal-college-of-anaesthetists-educational-campaign-on-capnography-during-cpr/)
13. Kodali BS. Capnography is vital during sedation. Capnography.com [Link](https://www.capnography.com/capnography-is-vital-during-sedation/)
14. Sammartino M, Volpe B, Sbaraglia F, Garra R, D'Addessi A. Capnography and the bispectral index: their role in pediatric sedation, a brief review. Int J Pediatr. 2010;2010:828347. doi:10.1155/2010/828347 [Link](https://pmc.ncbi.nlm.nih.gov/articles/PMC2952943/)
15. Saunders R, Struys MMRF, Pollock RF, Mestek M, Lightdale JR. Patient safety during procedural sedation using capnography monitoring: a systematic review and meta-analysis. BMJ Open. 2017;7:e013402. doi:10.1136/bmjopen-2016-013402 [Link](https://pmc.ncbi.nlm.nih.gov/articles/PMC5734204/)
16. Rosenberg H, Pollock N, Schiemann A, Bulger T, Stowell K. Malignant hyperthermia: a review. Orphanet J Rare Dis. 2015;10:93. doi:10.1186/s13023-015-0310-1 [Link](https://pmc.ncbi.nlm.nih.gov/articles/PMC4524368/)
17. Kaur A, Goel A, Dias R, Kesarkar N. An unusual cause of intraoperative hypercarbia. Indian J Anaesth. 2021;65(3):256–257. doi:10.4103/ija.ija\_1101\_20 [Link](https://pmc.ncbi.nlm.nih.gov/articles/PMC7989475/)
18. Damam S, Meshram RJ, Taksande A, et al. Navigating pediatric capnography: a comprehensive review of scope and limitations. Cureus. 2024;16(1):e53289. doi:10.7759/cureus.53289 [Link](https://pmc.ncbi.nlm.nih.gov/articles/PMC10905056/)